Your Reliable Partner for THC Measurements

Methane-Total Hydrocarbone Analyzer RS200

Applications

  • Emission Monitoring
  • THC / Methane Measurements
  • Optimizing Processes

Features

  • Heated Analytical Compartment (up to 200 °C / 473 °K)
  • Automatic Switch-Over Methane / THC
  • Auto-Range/Manual Range
  • Flame Out Indication
  • Automatic Hydrogen Cut-Off
  • Self-Monitoring Capabilities
  • Option: Sample Lines
  • Option: Filters
  • Option: Internal Zero-Gas / Combustion-Air Supply
  • Option: USB-Interface
Methane-Total Hydrocarbone Analyzer RS200

Specifications

(Reference Quantity : Full Range Deflection = FRD))

Measurement Ranges (Standard) 0-10 ppm Methane
0-100 ppm Methane
0-1.000 ppm Methane
Linearity +/- 2 % FRD (up to range 3)
Cross Sensitivity
(gaseous compounds
SO2,CO,CO2,NO,HCL,H2O)
< 4 % FRD
Limitof Detection < 2 % FRD
Zero-Point Drift < 2 % FRD/week
Sensitivity Drift < 4 % FRD/week
Response Time (T90) approx. 5 s
Sample Flow  
Signal Output 0-10 VDC or 0/4-20 mA
Range Identification DC Voltage
Status-ID GKW/Methane 1V / 2V
Relay Contacts
  • Concentration Pre-Alarm (Option)
  • Concentration Main-Alarm (Option)
  • Range ID
  • Flame Monitoring
  • Status (Calibration, Zerogas, Cal. Gas)
  • Alarms (Pressure, Temperature, Flame)
Dimensions 19" x 267 x 500mm (WxHxD)
Weight approx. 26,5 kg
Ambient Temperature + 5 ... + 35 °C
Main Power 230 V / 50 Hz or 115 V / 60(50) Hz / 800 Watts