Total Hydrocarbon Analyzer RS55-T/2
Applications
- Measurements at two Sample Points
- Emission Monitoring
- Optimizing Processes
- Combustion Control
- Threshold Limit Values (TLV) - Measurements

Features
- Heated Analytical Compartment (up to 160 °C / 320 °F)
- 2 Sample Inputs, Manual or Auto-Control
- Continuous / Discontinuous Measurement Mode
- Flame Out Indication
- Auto-Range / Manual Range
- Automatic Hydrogen Cut-Off
- Self-Monitoring Capabilities (Temperatures, Pressures, Flame)
- Option: Sample Lines
- Option: Filters
- Option: Internal Zero-Gas and Combustion-Air Supply
- Option: USB-Interface

Specifications
(Reference Quantity : Full Range Deflection = FRD)
| Ranges |
|
|
Linearity Sample Input Select |
+/- 2 % FRD (up to Range 4) Manual or Automatic, Time Base 1-99 Minutes |
| Oxygen Cross Sensitivity | < +/- 2 % FRD |
| Availability |
> 95 % |
| Limit of Detection | < 2 % FRD |
| Zero Drift | < 2 % FRD/week |
| Span Drift | < 4 % FRD/week |
| Response Time | approx. 1-2 s (without sample line) |
| Sample Flow Rate | approx. 220 l/h |
| Signal Output | 0-10 V or 0/4-20 mA |
| Status Contacts |
|
| Dimensions | 19" x 267 x 500 mm (WxHxD) |
| Weight | approx. 26,5 kg |
| Ambient Temperature | + 5 ... + 35 °C |
| Main Power |
230 V / 50 Hz or 115 V / 60(50) Hz / 800 Watts |